There have been no reports of patient injury or death, and there is no risk to patients who previously underwent a thrombectomy procedure with the Fetch 2 catheter. All reports of shaft breakage happened during the procedure, and the broken section was either removed while still partially attached to the catheter shaft or retrieved with a snare, without further patient complications. While unreported, the most severe potential outcome of this breakage is embolism of device fragments, which could lead to obstruction of blood flow or additional intervention to remove a device fragment surgically.
As part of the recall, all affected healthcare facilities were advised to discontinue use of all Fetch 2 catheters immediately and return unused product to Boston Scientific. Because Boston Scientific acquired the Fetch 2 catheter product line from Bayer Medical Care Inc., all recalled inventory is packaged and labeled as Bayer product. This device was manufactured between June 11, 2014 and February 19, 2016. There are currently 21,155 devices on the market subject to this recall.
Fetch 2 Catheter UPNs
FETCH2 US 109400-001
FETCH2 OUS 109400-002
FETCH2 Canada 109400-003
FETCH2 Japan 109400-004
FETCH2 EU 109400-005
To get more Information, please visit Boston Scientific website by clicking here.